A research team from the Anhui Institute of Optics and Fine Mechanics (AIOFM), the Hefei Institutes of Physical Science of the Chinese Academy of Sciences (CAS), in collaboration with the Hefei Cancer Hospital of CAS, has developed a new method to accurately measure hemoglobin levels in blood using near-infrared spectroscopy (NIRS).
Their study, published in Biomedical Signal Processing and Control, demonstrates how this method can greatly improve the reliability and accuracy of blood analysis with NIRS.
NIRS is valued for its convenience, non-invasiveness, and sensitivity. However, in whole-blood testing, the weak absorption features of hemoglobin (Hb) are often masked by strong water absorption, sample scattering, and instrument noise, which limits quantitative accuracy.
To overcome this problem, the team created an integrated processing and feature retrieval method that combines Wavelet Packet–Fuzzy Shrinkage denoising (WPT-FS) with the Whale Optimization Algorithm.
"We use 'fuzzy shrinkage' technique to remove noise from the signal, and whale optimization algorithm to find the key parts of the signal that show hemoglobin," explained Prof. HAN Xin, who led the team.
This approach first applies an adaptive "fuzzy shrinkage" threshold function to remove noise across multiple scales while preserving meaningful spectral information. Then, using swarm intelligence, the algorithm searches for optimal feature node combinations in the high-dimensional node space to reconstruct Hb-rich spectral features. Finally, Partial Least Squares (PLS) modeling is used to complete the quantitative analysis.
Tested on 106 whole-blood samples, the method significantly reduced background noise—from 58.68×10⁻⁵ in raw spectra to 1.8542×10⁻⁵—and achieved a root mean square error of prediction (RMSEP) of 2.0409 and a prediction coefficient of determination (R²_P) of 0.9746.
Further data augmentation experiments also confirmed the model’s robustness under more heterogeneous data distributions, achieving an average percentage error of about 1.7%.
According to the team, this work provides a reliable tool for non-invasive and accurate blood analysis, and could support future medical diagnostics and health monitoring.